Answer
This article covers suggestions to improve liquid pipetting. See also : Suggestions for better workflows
Suggestions
-
Aspirate slow and dispense fast in general.
Aspirate < 200 μl /s and dispense >250 μl /s
This prevents the formation of droplets on the tip after dispense.
Viscous liquids should be aspirated & dispensed slowly with a longer dwell time and a contact with the labware or liquid surface.
The pipetting speed should be the same as the volume aspirated but never exceeding the maximum flow rate for SQWERTY: 450 μl/s .
Samples should be free of any debris, foam/bubbles and droplets on the tube/well walls – lightly centrifuging will solve these problems
Don’t overfill the container as it may cause it to overflow during stage movement or pipette tip insertion.
-
Decreasing the aspiration flow rate will increase precision/accuracy.
10-70 μl /s is recommended for volumes less than 30 μl or with high viscosity or volatile fluids.
The faster the aspiration flow rate, the stronger the pull which could result in bubbles within the tip
Use an aspiration dwell time of at least 300 ms especially with 1000 μl tips. Higher viscosity liquids require further delays.
The submerged depth should be as small as possible 1-2 mm if possible but no more than 1 cm.
-
If using small volumes then add additional volume as a blow out / air gap / reverse pipetting.
When using a 250 μl pipette but transferring only 10 μl , add an additional 20 μl of volume with air or fluid.
Liquids like detergents/ buffers should be dispensed slowly to avoid foaming
Volumes lower than 20 μl need high dispense speed and a bigger air gap values to allow for it.
Large volumes should be dispensed at a slower speed.
0 mm dispense for greater workflow speed with slightly less accuracy. Touch transfer or surface level dispense for lower throughput but higher accuracy.
-
Air gap + Blow out should be less than 30 μl as there is no advantage to performance but the accuracy and precision may suffer.
The blowout should be bigger than the air gap and at least 5 μl to truly be effective.
Liquids with high vapour pressure need to be pre-wet to avoid liquid loss through evaporation. Do this up to 3 times but don’t spend too long as the delays could cause further liquid loss from the source via evaporation.
If you require more information please contact technicalsupport@singerinstruments.com for assistance.
Comments
0 comments
Please sign in to leave a comment.