Does the ROTOR(+) have inbuilt air filters? eg. Hepa Filter
Answer : The ROTOR does not come with an inbuilt air filtering system.As standard, the ROTOR is fitted with UV sterilisation and a protective screen. Although not sealed this does prevent most airb...
-
What are the recommendations for a mains air supply for the ROTOR(+)?
IntroductionIt is possible to run the ROTOR(+) from a Mains Air Supply. This may be beneficial for some customers who already have a supply in their lab. As well as the convenience and reduction in...
-
How to prepare for the arrival of the ROTOR(+)
IntroductionThis article includes all the necessary information required when preparing for your ROTOR(+) installation. A PDF document has been added to this article which includes this information...
-
How to perform Random Colony Screening Protocol on the ROTOR(+)
Answer: To pick randomly plated colonies into 96 well plates and screen for the phenotype in liquid culture it would be beneficial to include a preliminary screening step on the agar plates. The st...
-
Can the ROTOR(+) use compounds?
Answer: It is possible to use compounds in the ROTOR. Here are some examples of questions related to using compounds in the ROTOR(+).Questions: Has anyone used your pads to apply compounds dissolve...
-
What air pressure does the ROTOR(+) require?
Answer : The ROTOR(+) requires a minimum of 6 Bar - 0.6 MPa - 87 psi. The ROTOR has its own pressure valve which is set at 4.5 Bar - 0.45 MPa - 65 psi , but the supply pressure must be able to main...
-
What is the recommended agar volume for the Plus Plates used in the ROTOR(+)?
Answer: When using standard Plus Plates we recommend a minimum agar volume of 40ml.Although this is the recommended volume the ROTOR(+) will happily pin at any height of agar. The most important th...
-
What is the density range of the ROTOR(+)?
Answer: The ROTOR(+) is able to pin at all densities from 96 to 6144 and above. There is no software change required unless you have a very old software version. Please contact Singer Instruments ...
-
What is the recommended pinning density for the ROTOR(+)?
Answer: The ROTOR(+) is capable of pinning at 96, 384, 1536, 6144 densities on solid media and 96 and 384 in liquid media using our specially developed RePads.. Liquid: 384 has become the preferred...
-
What is the humming / buzzing noise coming from the ROTOR(+) during pinning?
Answer: This is perfectly normal and it is unlikely that there is a fault with the ROTOR(+). Please see the video below for an example of the noise during normal operation.What causes it?The noise ...