How to use the Software 'Program Options' in the ROTOR(+) software (Video)
Introduction : The ROTOR software for both the ROTOR HDA and ROTOR+ is very easy to use. The following videos demonstrate the Software Options for each version of the ROTOR. Please Note : These vi...
-
Why do the main circuit breakers trip when the ROTOR is turned on?
Answer : Due to the filters and motors fitted to the ROTOR there is likely to be current flow to earth in the ROTOR drive system. The EMC filters and the shielding on the cables are intentionally d...
-
What is the heat output of the ROTOR HDA and its compressor in BTU/hour?
Answer : The approximate BTU/hour values for the ROTOR HDA and Compressor are : ROTOR HDA : 147 BTU/hr* Compressor : 337 BTU/hr* * BTU - British Thermal Units Note: This article only applies to...
-
SporePlay User Guide
User Guide : The User Guide for the original SporePlay is attached to this article. Note: If you have the newer model SporePlay+ please follow the link. If you require more information please conta...
-
Does temperature change affect liquid handling?
Answer Any changes in temperature affects the fluid dynamics and thus the optimal liquid handling technique. If a heating or cooling step is required, outside of SQWERTY, let the liquids and labwar...
-
What is the volume transferred when pinning from liquid to agar? - PIXL
Answer: When transferring from liquid to agar on the PIXL the volume transferred is approximately 226nl. These results are based on pinning from a Corning™ Costar™ 96-Well, Cell Culture-Treated, Fl...
-
How to use the Manual Operation Feature in the ROTOR Software (Video)
Introduction : The video below demonstrates a software walk-through of how to create a basic custom routine using the Manual mode on the ROTOR (supplied before July 2021). Note: To identify the d...
-
Why do I need to select the right PhenoBooth for my software?
Answer : The PhenoBooth and PhenoBooth+ have different cameras, and therefore the software needs to know which one to connect to. Check that the correct PhenoBooth is selected here from the top men...
-
How to resolve main supply voltage variations for the ROTOR.
Answer : Some customers in the USA reported issues with their ROTOR which was caused by the variations in their main supply voltage. This can be due to having a large item such as a fridge on the s...
-
If we install the slide holder can we take a picture?
Answer : Yes you can take a picture using the MSM 400 software. If you require more information please contact technicalsupport@singerinstruments.com for assistance.