Articles
Recent activity by Ian-
Can we re-use PlusPlates and RePads?
Answer: Singer PlusPlates and RePads are not recommended for re-use. We supply them in sterile packaging intended for one use only to reduce any possibilities of contamination.Reasons: Sterility is...
-
Can I autoclave the ROTOR(+) Dump Drawer and Pad Hopper?
Answer : YesThe ROTOR Dump Drawer and the Pad Hopper, including the Pad Support block can be autoclaved.Dump Drawer Pad Hopper with Support BlockIf you require more information please contact techn...
-
How successful is the 6144 density on the ROTOR(+)?
Answer:There are a few Singer Instruments customers who have been successfully pinning at 6144 density with the ROTOR.Laboratories that use this method grow the 6144-density plate for a shorter tim...
-
Is there a risk analysis available to warn employees of the dangers posed by the ROTOR(+)?
IntroductionRisk Analysis for the ROTOR(+) should be undertaken by the customer's own Health and Safety Department based on the following information.ROTOR HDAThe ROTOR has many safety protections ...
-
Where to purchase a replacement UV Tube for the ROTOR(+)
Answer: Replacement UV tubes are no longer kept in stock by Singer Instruments. They can be purchased from Amazon here : G30T8 - 30 watt T8 Ultraviolet UV Tube 36" Length Medium Bi-pin Base or eBay...
-
Where can I find the Compressor Manuals for the ROTOR(+)?
Answer : Since the release of the ROTOR in 2005/6 we have supplied 5 different compressors. Manuals for these compressors are left with the ROTOR during commissioning but we appreciate that they ar...
-
What fuses are used in the ROTOR?
Answer : The table below lists the fuses used in the ROTOR please contact Singer Instruments for fitting instructions. A selection of these fuses are included in the ROTOR toolkit supplied with the...
-
Can the computer (MCI) be fitted to either end of the ROTOR(+)?
Answer : This depends on the ROTOR version.ROTOR HDAThe ROTOR touch-screen computer and even the older ROTOR's with laptop shelves, can be fitted to either end of the ROTOR.It requires swapping the...
-
Can I use deep well plates (eg. 2.2 ml) in ROTOR(+)?
Answer : It is not possible to use 2.2ml deep well plates in the ROTOR (pre-2021). This is due to height restrictions inside the ROTOR. The ROTOR+ will accept the deep well plates. It still uses th...
-
Can I use low profile multi well plates on ROTOR(+)?
Answer : These plates are not high enough to enable the ROTOR(+) to be able to operate normally however there is no reason why they can not be used if required.The gripper head which holds the pads...